Hi GC9020
1. System capability
- Assembly simultaneously: 2 inlets+3 detectors(FID, TCD, ECD, FPD and NPD)
- Automatic control can be done from local keyboard and networked PC
- Carrier gas control: EPC/AFC
- 8 channels independent temperature control
- Automatic liquid sampling available
2. EPC/AFC gas system
- Carrier and makeup gas settings selectable for He, H2, N2 and Ar
- Psi, KPa, Bar units selectable
- Pressure control range: 1~600KPa, Pressure accuracy: 0.01Psi, Programmable pressure ramp RSD ≤0.5%
- Flow rate control range: 0~600ml/min, flow rate accuracy: 0.1ml/min, flow rate RSD≤0.1%
- Programmable pressure/flow rate ramping: 8 steps
- Flow sensor accuracy: ≤±3%, detector module accuracy: ≤±7%
3. Column oven
- Column oven dimension: 278×310×165mm=14.2L, accommodate up to 2pcs 105m×0.53mm ID capillary column
- Temp. control range: RT+5℃~450℃(0.1℃ increment)
- Programming temp.-ramp : any step
- Temp. ramp speed: 0.1~120℃≤1 minute
- Fast cool down speed: 350~100℃≤3.5 minutes
- Temp. accuracy: ±0.1℃
- Max run time: 999.99 minutes
4. Inlets
- Packed purge injection port(PPIP)
- Split/splitless capillary port(S/SL)
- Max temperature: 450℃
- Split ratio: 4500:1
- Gas saver mode to reduce gas consumption without compromising performance
- Programmable temp./pressure ramp: 8 steps
- Valve injection is available
5. Workstation and data communications
- LAN
- Windows xp/7 support
- Signal capture from 2 detectors simultaneously
- Sampling frequency: 100Hz
- Sampling speed: 50 times/sec
- Unlimited peak quantity
- Self-diagnosis: intelligent automatic error identification and self-protection
- Compatible with Clarity to fully support FDA-21CFR PART 11, SST AND IQ/OQ (optional)
6. Basic info
- Voltage: 220V±10%, 50Hz
- Power: 3000W
- Dimension: 645×500×555mm
- Net weight: 55KG
- Votage: 220V±10%, 50Hz
7. Detectors